In order to assure optimum pressure drop across the splitter, please specify the actual prep input flow and the targeted split flow range under Custom Specification field. Please contact with us for ordering information.

This product is not returnable since it's a custom Prep Inlet Flow/Split Flow Range.

 

Note:

To check if the low flow capillary resistor is clogged, you can disconnect the capillary resistor from the make up inlet tee, run the splitter at approximately 1,000 PSI, and look for flow out of loose end of capillary.  You will need to give it some time to accumulate.  If the low flow capillary resistor is clogged, you will need to order Adjustable Makeup-Flow Splitter, Low Flow Capillary Resistor, ASI P/N: 690-POCR https://www.hplc-asi.com/adjustable-makeup-flow-splitter-low-flow-capillary-resistor/.

Current stock: 0

Custom Prep Inlet Flow, Adjustable Makeup-Flow Splitter FV

US$3,706.00
Part Number: 691-POCS-CSFV
Prep Inlet Flow *
Split Flow Range *
Specifications:

Port size: 10-32 CPI (standard HPLC fittings for 1/16" tube)

Compatible with Thermo Scientific™ nanoViper™ Fingertight Fittings

Maximum Operating Pressure: 350 bar (5,000 PSI)

Part Number:
691-POCS-CSFV
Wetted Material:
Stainless Steel, PEEK, Teflon, Fused Silica