













|
 |
HyperShear™
Static Mixer
Application Notes
Selecting the Right Size Mixer Cartridge
ASI offers static mixers with volumes that range from
2 micro liters to 1,000 micro liters. Choosing the right size is a trade
off between delay volume, mixing noise, and step gradient definition
and repeatability.
The following observations will provide some guidelines
to help choose the right size mixer.
-
For any given flow rate, the more the mixing
volume the better the mixing, and the lower the baseline noise.
-
The smaller the mixing volume, the better the
definition and sharpness of step gradients.
-
Multi-pump gradient systems typically require
far less mixing volume than low pressure single pump gradient systems.
-
A 150 µL in-line static mixer can be added after
a dynamic mixer to further reduce mixing noise.
Multi-pump High Pressure Gradient Systems
- Linear Gradients
-
Larger mixing volume can be tolerated for a particular
flow rate, with the larger the volume, the lower the mixing noise.
The upper limits to mixing volume will be the maximum delay time
that can be tolerated, and possible distortion (tailing) of the
gradient at the beginning and end of the gradient. The lower limit
will be defined by the amount of mixing noise that can be tolerated.
-
See table 1 for
specific recommendations.
- Binary or Ternary _ Steady State Composition
-
Always select the largest volume that will still
provide an acceptable delay volume. In general, the more mixing
volume, the better the mixing will be. For most pump systems, a
150 µL cartridge will provide adequate mixing.
| Examples of this type of pump system include: |
Shimadzu 10 AD and 10 ADVP
Beckman System Gold®
Gilson® Model 305
Hewlett Packard® Model 1100 binary |
Single-pump Low Pressure Mixing Gradient Systems
- Linear Gradients
These systems generally require more mixing volume
to perform linear gradients than multi pump high pressure systems.
The following will explain why this is the case. In a low pressure
system the composition can only be changed once every pump stroke.
Since the pump stroke volume of most pumps is 100 µL, and it takes
a mixer volume that is about 3 times the batch volume to provide
adequate mixing, we need 350 µL of mixer volume, at least, to do
adequate mixing. More insoluble combinations may require even more
mixing volume.
In general, choose the largest size mixer cartridge
that will still provide an acceptable delay volume. For most applications
this will be at least 350 µL.
Binary or Ternary - Steady State Composition
Always select the largest volume that will still
provide an acceptable delay time. In general, the more mixing volume,
the better the mixing will be. For most applications this will be
at least 350 µL.
| Examples of this type of pump system include: |
Hewlett Packard® 1090
Perkin Elmer series 200
TSP Spectra Vision®
Waters model 626
Varian Star® 9000 |
Mixer Cartridge Seclection Guide for High Pressure Mixing
Linear Gradients - High Pressure Mixing
-
Larger mixing volume can be tolerated for a particular
flow rate, with the larger the volume the lower the mixing noise.
The upper limits to mixing volume will be the maximum delay time
that can be tolerated, and possible distortion (tailing) of the
gradient at the beginning and end of the gradient. The lower limit
will be defined by the amount of mixing noise that can be tolerated.
The following cartridge volumes are a compromise between these two
limits.
|
Flow
|
Mixer Cartridge Volume
|
|
0-5
micro liter/min. |
5
micro liter |
| 5-10 |
10 |
|
10-20 |
25 |
| 20-150 |
50 |
|
150-500 |
150 |
| 500+ |
250 |
|