QuickSplit™ Adjustable Makeup-Flow Splitter

 Adjustable Makeup-Flow Splitter Performance

The chromatographic data below compares the dispersion (band broadening), measured as variance, at 10 μL/min. between an ideal system with direct flow (Figure 19) and a split system incorporating the ASI QuickSplit™ Adjustable Makeup-Flow Splitter (Figure 20). Split flow variance is measured with and without the addition of a small amount of makeup flow. The data proves conclusively that ASI splitter dead volume does not contribute significantly to overall system dispersion. It also demonstrates the advantages of adding a makeup flow not only to improve peak shape but acts as an additional tool to optimize the timing sequence between Mass Spectrometer detector and the fraction collector.

Conditions: Variance Calculation:
HPLC System: Shimadzu LC10 AD VP
Detector: UV @254 on column
Solvent: Water
Injection vol: direct 600nL, split 135µL
Variance = Sigma2 = (Wh x F)2 = µL2
W = Peak width at half height
F = Flow rate
Figure 19 Figure 20